AMPEROMETRIC CELL – CAC-HT-B
MANUFACTURER: STEIEL ELETTRONICA Srl
CAT NO: CAC-HT-B
- Description
- Size Guide
- SPECIFICATONS
- DOWNLOADS
- Vendor Info
- More Products
Description
AMPEROMETRIC CELL – CAC-HT-B
DESCRIPTION OF AMPEROMETRIC CELL – CAC-HT-B
Three-electrode amperometric sensor, suitable for the control of chlorine or chlorine dioxide concentration in hot water circuits and in dirty water. To be installed in a proper probe holder with regulation of inlet water flow and proximity sensor.
TECHNICAL FEATURES
- Three-electrode amperometric probe, with protective cap and specific electrolyte, which ensures stable and protected measurement conditions
- Measurement range: 0.02 … 2 ppm (chlorine / chlorine dioxide)
- Continuous and reliable monitoring, for an efficient management of the processes
- Thermo-compensated readings, thanks to the inbuilt temperature sensor
- Working conditions: temperature max 70 °C; pressure max 8 bar; recommended flow rate 60 l/h; conductivity of the sample 50 … 10’000 µS/cm
- PEEK probe body and electrode protection cap
- Shielded, 4-wire cable
| CELL | ANALYSIS OF | RANGE (ppm) |
RECOMMENDED USE | TEMPERATURE / PRESSURE MAX |
CONNECTION |
| CAC-HT-B | Chlorine / ClO2 |
0.02-2.0 | Disinfection of sanitary hot water; analysis in dirty water |
70 °C / 8 bar | Cable 1 m
Cable 1 m + M8 |
- Three-electrode amperometric probe, with protective cap and specific electrolyte, which ensures stable and protected measurement conditions
- Measurement range: 0.02 … 2 ppm (chlorine / chlorine dioxide)
- Continuous and reliable monitoring, for an efficient management of the processes
- Thermo-compensated readings, thanks to the inbuilt temperature sensor
- Working conditions: temperature max 70 °C; pressure max 8 bar; recommended flow rate 60 l/h; conductivity of the sample 50 … 10’000 µS/cm
- PEEK probe body and electrode protection cap
- Shielded, 4-wire cable
| CELL | ANALYSIS OF | RANGE (ppm) |
RECOMMENDED USE | TEMPERATURE / PRESSURE MAX |
CONNECTION |
| CAC-HT-B | Chlorine / ClO2 |
0.02-2.0 | Disinfection of sanitary hot water; analysis in dirty water |
70 °C / 8 bar | Cable 1 m
Cable 1 m + M8 |
Vendor Information
-
-
Chemicals, Laboratory Reagents & Standards, Chemicals, Laboratory Chemicals
Hyamine® 1622 solution (0.004 M) for volumetric analysis
$0.00Select options This product has multiple variants. The options may be chosen on the product pageQuick View