Benchtop pH/Conductivity Meter Bante901 P-UK

$1,058.73

Benchtop pH/Conductivity Meter Bante901 P-UK

High-accuracy multiparameter water quality meter contains 12 options, the meter is suitable for measuring the pH, mV, conductivity and total dissolved solids

SKU: EA-5165-117

Description

Benchtop pH/Conductivity Meter Bante901 P-UK

Benchtop pH/Conductivity Meter Bante901 P-UK

pH Mode:
• Multiparameter water quality meter is equipped with a large backlit LCD display.
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
• Automatic electrode diagnosis shows the pH slope and offset.

Conductivity/TDS Modes:
• 1 to 5 points calibration with automatic recognition for conductivity standards.
• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear
and pure water compensation modes.

pH Conductivity Meter

Other Features:
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-Read function senses and locks the measurement endpoint.
• Calibration due alarm prompts user to calibrate the meter regularly.
• Calibration report provides the details of electrode performance.
• Setup menu allows setting the number of calibration points, stability criteria, temperature
unit, date and time, auto-power off, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface for data transfer and timed interval readings.

ORDER INFORMATION:

 Order Code Accessories
   Bante901 P  – UK
P11 glass pH electrode
K10 conductivity electrode
TP-10K temperature probe
pH4.01/7.00/10.01 buffer sachets
Conductivity standards 84µS/cm, 1413µS/cm, 12.88mS/cm
USB cable
DC5V power adapter

SPECIFICATIONS:

Model Bante901
   pH
   Range -2.000~20.000pH
   Accuracy ±0.002pH
   Resolution 0.001, 0.01, 0.1pH, selectable
   Calibration Points 1 to 5 points
   pH Buffer Options USA, NIST, DIN or custom
   mV
   Range ±1999.9mV
   Accuracy ±0.2mV
   Resolution 0.1, 1mV, selectable
   Conductivity
   Range 0.01~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
   Accuracy ±0.5% F.S
   Resolution 0.001, 0.01, 0.1, 1, automatic
   Calibration Points 1 to 5 points
   Calibration Solutions 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
   Temperature Coefficient Linear (0.0~10.0%/°C) or pure water compensation
   Cell Constant K=0.1, 1, 10 or custom
   Reference Temperature 20°C or 25°C
   TDS
   Range 0~10.00, 100.0, 1000ppm, 10.00, 100.0ppt (Max. 200ppt)
   Accuracy ±1% F.S
   Resolution 0.01, 0.1, 1, automatic
   TDS Factor 0.1~1.0 (Default 0.5)
   Temperature
   Range 0~105°C, 32~221°F
   Accuracy ±0.5°C, ±0.9°F
   Resolution 0.1°C, 0.1°F
   Offset Calibration 1 point
   Calibration Range Measured value ±10°C
   General
   Temperature Compensation 0~100°C, 32~212°F, manual or automatic
   Stability Criteria Low or high
   Calibration Due Alarm 1 to 31 days or off
   Memory Stores up to 500 data sets
   Output USB communication interface
   Connector BNC, 6-pin DIN
   Display 6.5″ custom LCD
   Power DC5V, using AC adapter, 220V/50Hz
   Dimensions 210(L)×188(W)×60(H)mm
   Weight 1.5kg

Vendor Information