CS-iTEC Residual Oil Sensor S 120-P Connectable to S 551 +Transport Case
$8,898.29
Manufacturer: CS-iTEC
Product code: AME-5046-1
- Description
- Size Guide
- Vendor Info
- More Products
Description
CS-iTEC RESIDUAL OIL SENSOR S 120-P CONNECTABLE TO S 551 + TRANSPORT CASE
CS-iTEC Residual Oil Sensor S 120-P Connectable to S 551 +Transport Case
The new residual oil sensor S 120 monitors oil contents of compressed air and gases permanently or for spot checks when used as portable unit in conjunction with S 551. The simple installation and the outstanding performance makes S 120 the ideal choice when residual oil contents needs to be measured and monitored.
Oil free compressed air is not an easy task to be achieved. Monitoring is a must in many industries and applications to avoid contaminations in products and risks for health of humans. S 120 makes this monitoring task affordable and reliable.
For best accuracy and long term stability, the sensor applies an automatic calibration. Sensor contaminations and sensor life time are monitored and will be indicated to the user. An over range detection will remove the sampling air from the sensor to protect it against contamination.
Application
- Medical air
- Pharmaceuticals
- Breathable air
- Marine air
- Food and beverage
- Medical engineering
- High speed trains
- Semiconductor fabs
- Conveyance of hygroscopic food
- High tech processes
- Electronics industry
Limits of oil vapor
Compressed air class 1 (EN ISO 8573-1): 0.01 mg/m³
Medical applications (EAB 407/1238): 0.1 mg/m³
Breathing apparatus (EN 12021): 0.5 mg/m³
Product features
- Measures residual oil contents in compressed air and other gases
- Can be used for permanent or in portable applications
- Measures down to 0.003 mg/m³
- Easy connection through sampling hose and quick connect
- Output signals:
– 4 … 20 mA
– RS-485 / Modbus/RTU
– Relay switch (NO) - PID sensor for highest accuracy
- Service and Alarm indication through LED
- Connectable to CS-iTEC displays and data loggers as well as third parties displays and control units
Technical data
| Technical data S 120 | |
| Measuring range | 0.003 … 10.00 mg/m³ |
| Sensor type | PID (photoionization detector) |
| Detection limit | 0.003 mg/m³ |
| Accuracy | 5% of reading +- 0.003 mg/m³ |
| Operating pressure | 3 … 15 barg (higher pressure on request) |
| Gas humidity | < 40% rel. humidity, no condensation |
| Sample flow rate | < 2 l/min, measuring gas is released to ambient |
| Gas connection | 6 mm quick connect |
| Electrical connection | M12 connector |
| Sensor life time | 6000 operating hours. Will be indicated. Sensor exchange by service |
| Gas temperature | -20 … 50 ºC (at inlet) |
| Ambient conditions | -20 … 50 ºC |
| Transport temperature | -30 … +70 °C |
| Output signal | 4…20 mA (0 … 10 mg/m³) RS-485, Modbus/RTU Relay: NO, 32 VDC / 500 mA |
| Power supply | 24 VDC ± 5%, 8 W |
| Display & data logger | 5” touch screen, 100 million values (option) |
| Application | Downstream of activated carbon filters Downstream of oil-free compressors Wherever upstream drying and filtration is applied |
| Casing/dimensions | PC, Al alloy, 271 X 205 X 91 mm |
| Classification | IP65 |
| EMC | According to IEC 61326-1 |
| Settings | Various sensor settings can be performed through CS-iTEC display units or through the related service software |
| Weight | 2400 g |
| Sample rate | 1 s |