Insert HR for ADT878-1210 Thermocouple Calibration Furnace

Insert HR for ADT878-1210 Thermocouple Calibration Furnace

$1,403.00

MANUFACTURER: ADDITEL

MODEL NO:  ADT110-878-TC-INSERT-HR

Insert DR for ADT878-1210 Thermocouple Calibration Furnace

SKU: AME-41-486

Description

Insert HR for ADT878-1210 Thermocouple Calibration Furnace

Product Highlights

Thermocouple Calibration Reference Furnace, 100°C to 1210°C, Insert HR, 110V

  • Temperature control from 100°C to 1210°C
  • Two models to choose from: Reference (ADT878) and Standard (ADT875)
  • Display Accuracy of ±1.5°C (ADT878)
  • 4 on-board measurement channels (PC option)
  • Process calibrator option provides a multi-channel readout for TCs, switches and transmitters, including task documentation and HART communication
  • Portable, rugged and quick to temperature
  • Self-calibration feature (PC option)
  • Multi-zone temperature control
  • Internal and external sensor control (PC option)
  • Metallic interchangeable inserts
  • Wi-Fi and Bluetooth capable
  • Color touch screen display
  • ISO 17025-accredited calibration w/data included
  • Patent pending technology

We understand the many challenges associated with thermocouple calibration work. That is precisely why we decided to introduce the ADT875-1210 and ADT878-1210 Thermocouple Calibration Furnaces.

With an unmatched stability, uniformity and an optional on-board process calibrator, calibrating thermocouples has never been easier. With two separate units to choose from, the ADT875-1210 and ADT878-1210 furnaces include a patented multi-zone temperature control which provides a never before seen, highly stable and uniform heat source to ensure you get the best possible results from a modest investment. With metallic interchangeable inserts, users have the flexibility needed to service a wide variety of UUT

 

General Specifications
875-1210 878-1210
Temperature Range 100°C to 1210°C
Display Accuracy ±1.2°C @ 100°C ±1.0°C @ 100°C
±1.2°C @ 300°C ±1.0°C @ 300°C
±1.2°C @ 600°C ±1.0°C @ 600°C
±1.6°C @ 900°C ±1.2°C @ 900°C
±2.0°C @ 1210°C ±1.5°C @ 1210°C
Stability ±0.1°C
Axial Uniformity (20mm zone) ±0.6°C @ 100°C ±0.4°C @ 100°C
±0.9°C @ 300°C ±0.6°C @ 300°C
±1.2°C @ 600°C ±0.8°C @ 600°C
±1.5°C @ 900°C ±1°C @ 900°C
±1.5°C @ 1210°C ±1°C @ 1210°C
Radial Uniformity ±0.2°C @ 100°C ±0.2°C @ 100°C
±0.3°C @ 300°C ±0.3°C @ 300°C
±0.4°C @ 600°C ±0.4°C @ 600°C
±0.8°C @ 900°C ±0.6C @ 900°C
±1°C @ 1210°C ±0.8°C @ 1210°C
Loading Effect ±0.5°C
Environmental Conditions 8°C to 38°C guaranteed accuracy
0°C to 50°C, 0% to 90% RH non-condensing,
3000 M altitude for normal operation
Storage Conditions -20°C to 60°C
Immersion Depth XR style inserts = 138 mm (5.43″)
XS style inserts = 116 mm (4.57″)
(see insert ordering info for more details)
Insert Size – OD 24.8 mm (0.98 inches)
Heating Time 50 min: 23°C to 1210°C
Cooling Time 50 mins:1210°C to 300°C 50 mins: 300°C to 50°C 55 mins:1210°C to 300°C 55 mins: 300°C to 50°C
Typical Time to Stability 15 min
Resolution 0.01°C
Units °C, °F, and K
Display 6.5 in (165 mm) color touch screen
Size (H x W x D) 345 x 170 x 330 mm (13.6 x 6.7 x 13.0 in)
Weight 10.6 kg (23.4 lbs)
Power Requirements 115 V AC (±10%) or 230 V AC (±10%), switchable, 45-65 Hz, 580 W
Mechanical Testing Vibration: 2 g (10-500 Hz), 30 min for 2 sides
Impact: 4 g three times
Drop test: 500 mm (19.6 in)
Communication USB A, USB B, RJ45, WiFi, Bluetooth
Localization English, Chinese, Japanese, Russian, German
Warranty 1 year
Input Specifications (Process Calibrator [PC] Option)
875-1210 878-1210
TC Measurement Channels Patented TC terminals:
Accepting S, R, K, B, N, E, J, T, L, and U
TC Measurement Accuracy Type K Ch. 1-4 (excluding sensor) ±0.182°C @ 100°C ±0.172°C @ 100°C
±0.266°C @ 300°C ±0.236°C @ 300°C
±0.310°C @ 600°C ±0.251°C @ 600°C
±0.397°C @ 900°C ±0.304°C @ 900°C
±0.517°C @1210°C ±0.382°C @ 1210°C
TC Range €”75 mV to 75 mV (UUT Channels 1-4)
€”18 mV to 18 mV (Reference Channel)
TC Resolution 0.0001 mV, Input Impedance < 10Î
You've just added this product to the cart: