Description

Kit without cell, incl. Calibration solutions and cell holder

Kit without cell, incl. Calibration solutions and cell holder

Product Description of Kit without cell, incl. Calibration solutions and cell holder

  • Kit without cell, with standard solution 1413 µS and 12,88 mS
  • Cell holder and accessories
  •  Selectable cell constant allows for use with 2 or 4 plate cell
  • GLP and IQ-OQ manual available
  • Automatic and manual compensation of temperature with reference value 15…30 °C
Technical data

Cond. Measuring range
Resolution cond
Relative accuracy
TDS Measuring range
TDS factor
Relative Accuracy
Temperature range
Accuracy/Resolution Temp.
Temperature compensation
Temperature calibration
Calibration points
Calibration standards recognized

Indication of cell condition
GLP
CALDUE
Memory
Inputs
Protection class
Power supply

Dimensions
Weight

EC-40 cond

0,00 – 200
Automatic scale
±0,5 % value reading
0,1 mg/l – 100 g/l
0.40…1.00
±0,5 % value reading
0 – 100
0,1/±0,2
Manual or automatic
Yes
1 bis 4
84, 1413 µS / 12.88,
111.8 mS, benutzerdef.
Yes
No
No
No

IP 54
AC/DC adapter
9 V, 300 mA
200 x 220 x 100
950 g
EC-45 cond

0,00 – 1000
Automatic scale
±0,5 % value reading
0,1 mg/l – 500 g/l
0.40…1.00
±0,5 % value reading
-20 – 200
0,1/±0,2
Manual or automatic
Yes
1 bis 4
84, 1413 µS / 12.88,
111.8 mS, benutzerdef.
Yes
Yes, date/time
Yes
1000 Data
2 x USB
IP 54
AC/DC adapter
9 V, 300 mA
200 x 220 x 100
950 g