Temperature & Humidity Transducer, P18L
$108.67
Temperature & Humidity Transducer, P18L
Manufacturer: LUMEL
Product Code: P18L
- Description
- Size Guide
$108.67
Manufacturer: LUMEL
Product Code: P18L
Features:
The P18L transducer is a device destined for the continuous measurement and conversion of relative humidity and ambient temperature into a digital way and into a voltage or current standard signal. The transducer programming is possible through the RS-485 interface.
Applied sensor shields (table 1) enable the application of the P18 transducer to operate in various ambient conditions. The transducer is designed to be mounted on a wall.
P18L (with local display)
Basic parameters: |
|
range of relative humidity measurement (RH) |
0…100% without condensation1) |
basic error of humidity conversion |
± 2% of the range for RH = 10…90% |
hysteresis for humidity measurement |
± 1% |
range of temperature measurement (T) |
– 20…60ºC2) |
basic error of temperature conversion |
± 0.5% of the range |
calculated quantities |
absolute humidity (a) [g/m3] |
additional errors: |
± 25% of the basic error/10ºC |
RS-485 digital output: |
|
transmission protocol |
MODBUS |
baud rate |
4800, 9600, 19200, 38400, 57600 bit/s |
mode RTU: |
8N2, 8E1, 8O1, 8N1 |
maximal response time |
300 ms |
Analog output (optional): |
|
current |
4…20 mA |
voltage |
0…10 V |
maximal load capacity of the current output |
100 Ω |
minimal load capacity of the voltage output |
1k Ω |
Rated operating conditions: |
|
supply |
9…24 V a.c./d.c. |
consumption |
< 0.5 VA |
ambient temperature |
– 30…23…85ºC |
relative air humidity |
< 95% 3) |
rate of air flow |
≥ 0.5 m/s4) |
preheating time |
15 minutes |
protection degree ensured by the housing |
IP 65 |
weight |
125g |
fixing |
on a wall |
working position: |
– in applications non-exposed to a direct contact with water: any |
Electromagnetic compatibility: |
|
noise immunity |
acc. to EN 61000 -6-2 |
noise emission |
acc. to EN 61000 -6-4 |
Security requirements acc. to EN 61010-1 |
|
installation category |
III |
pollution level |
2 |
phase-to-earth working voltage |
50 V |