Basic parameters:
|
|
range of relative humidity measurement (RH)
|
0…100% without condensation1)
|
|
basic error of humidity conversion
|
± 2% of the range for RH = 10…90%
± 3% for the remaining range
|
|
hysteresis for humidity measurement
|
± 1%
|
|
range of temperature measurement (T)
|
– 20…60ºC2)
|
|
basic error of temperature conversion
|
± 0.5% of the range
|
|
calculated quantities
|
absolute humidity (a) [g/m3]
dew-point temperature (Td) [ºC]
|
|
additional errors:
temperature influence
|
± 25% of the basic error/10ºC
|
|
RS-485 digital output:
|
|
transmission protocol
|
MODBUS
|
|
baud rate
|
4800, 9600, 19200, 38400, 57600 bit/s
|
|
mode RTU:
|
8N2, 8E1, 8O1, 8N1
|
|
maximal response time
|
300 ms
|
|
Analog output (optional):
|
|
current
|
4…20 mA
|
|
voltage
|
0…10 V
|
|
maximal load capacity of the current output
|
100 Ω
|
|
minimal load capacity of the voltage output
|
1k Ω
|
|
Rated operating conditions:
|
|
supply
|
9…24 V a.c./d.c.
|
|
consumption
|
< 0.5 VA
|
|
ambient temperature
|
– 30…23…85ºC
|
|
relative air humidity
|
< 95% 3)
|
|
rate of air flow
|
≥ 0.5 m/s4)
|
|
preheating time
|
15 minutes
|
|
protection degree ensured by the housing
|
IP 65
|
|
weight
|
125g
|
|
fixing
|
on a wall
|
|
working position:
|
– in applications non-exposed to a direct contact with water: any
– in applications exposed to a direct contact with water: with the
sensor chamber directed towards the ground.
|
|
Electromagnetic compatibility:
|
|
noise immunity
|
acc. to EN 61000 -6-2
|
|
noise emission
|
acc. to EN 61000 -6-4
|
|
Security requirements acc. to EN 61010-1
|
|
installation category
|
III
|
|
pollution level
|
2
|
|
phase-to-earth working voltage
|
50 V
|