Thermoreactor, COD Analysis – ECO16
$1,011.55
ECO16 Thermoreactor 115V/50-60Hz
Manufacturer: Velp
Product Code: EA-5726-2
16-position thermoreactor for Chemical Oxygen Demand (COD) with settable temperature (up to 160 °C) and operating time (up to 199 minutes).
- Description
- Size Guide
Description
Description
Accurate Operation
- Electronic temperature control
- Temperature regulation from ambient to 200 °C
- Analysis time can be set from 1 to 199 minutes or continuous
Large Sample Numbers
- Processes up to 14 x Ø 16 mm ml test tubes
- Plus 2 x Ø 22 mm test tubes simultaneously
- LCD countdown display shows the temperature and remaining time
Reduced Running Cost
- COD analysis in only 30 minutes at 160 °C
- 16 mm diameter tubes minimize reagent consumption
Safety in Use
- Automatic shut-down at run end
- Acoustic signals, e.g., at shut-down
- Antisplash cover protects the user
GENERAL FEATURES | |
Construction material: | Epoxy painted metal structure |
Holes number and diameter: | 14 pos. (Ø 16 mm) and 2 pos. (Ø 22 mm) |
Set temperature: | display visualization |
Countdown: | display visualization |
End of the cycle: | acoustic signal with automatic switch-off |
Power: | 550 W |
Weight: | 3,8 Kg (8.4 Ib) |
Dimensions (WxHxD): | 168x110x269 mm (6.4×4.3×10.6 in) |
PERFORMANCE | |
Selectable working temperatures: | from room temperature to 160 °C, resolution 1 °C |
Selectable working times: | from 0 to 199 minutes or continuous operation |
HEATING BLOCK TECHNICAL DATA | |
Temperature stability: | ± 0.5 °C |
Temperature homogeneity: | ± 0.5 °C |
Temperature precision: | ± 1 °C |
Overtemperature safety | |
SIGNALS | |
Reaching of set temperature: | visual signal |
Countdown: | visual signal |
End of cycle | acoustic and visual signal |
Probe interruption: | acoustic and visual signal |
Broken probe: | acoustic and visual signal |
Exceed temperature range: | acoustic and visual signal |