WALL INSTRUMENT FOR SINGLE MEASUREMENT – S507-ACL
WALL INSTRUMENT FOR SINGLE MEASUREMENT – S507-ACL
MANUFACTURER: STEIEL ELETTRONICA Srl
CAT NO: S507-ACL
- Description
- Size Guide
- SPECIFICATONS
- DOWNLOADS
- Vendor Info
- More Products
Description
WALL INSTRUMENT FOR SINGLE MEASUREMENT – S507-ACL
Input for CLE11 (or CLE12) cell; range to be specified
S507-ACL is a series of single-measurement instruments designed for wall installation, and suitable for use in the harsh industrial environments thanks to its watertight case.
The series includes different models, for measuring pH, ORP, conductivity, resistivity, free chlorine, total chlorine, bromine, chlorine dioxide, hydrogen peroxide, ozone, peracetic acid, turbidity and a version with mA input.
This instrument is the ideal solution for monitoring water in industrial processes, purification and conditioning, in the agri-food and livestock industry.
TECHNICAL DATA
- Input for one measurement sensor
- Input for Pt100 temperature sensor, for measurement and thermo-compensation
- Power supply 230 V~ ±10% 50/60 Hz (115 or 24 V~ or 24 V- on request)
- Two relay outputs configurable as intervention threshold or alarm
- “Dosing-shot” output option to control an electronic-free dosing pump
- Current output, fully configurable and galvanically isolated
- Serial port for communication with PC, data recorder, RW14
- Digital input to disable all outputs
- Multi-language menu (Italian, English, French, Spanish)
- Watertight housing
Specification
- Power Supply: 230 V~ ±10% 50/60 Hz (115 or 24 V~, or 24 V- upon request)
- Power Consumption: 7 VA max (models with relay output)
70 VA (models with K1 configured as “dosing-shot”) - Protections: PTC fuse appropriate to the power supply voltage.
If K1=dosing-shot and 230 V~ power supply, F 250 mA, 5×20 fuse - Environment: storage temperature -20 to +60 °C
operating temperature -10 to +50 °C
RH max 90% noncondensing - Display: 2-row, alphanumeric LCD (16 digits per row), with backlight.
- Digital Inputs: 1 OFF/FLW input for disabling outputs; for voltage-free contact; 5 V voltage at terminals, current max 5 mA.
1 AUX (auxiliary) input, available on special models.
1 TC (temperature control) input, internal input, only for K1= dosing-shot - Dosing-shot Output: factory setting of K1 relay for directly driving of the pump magnet; standard pulse 300 V and 90 msec, max frequency 120 pulses/minute (not available for 24 V- or 24 V~ power supply versions)
- Current Output: 0-20 or 4-20 mA (selectable through software), max load 600 Ohm, error max 0.5% FS; provided with galvanic insulation.
- Serial Output: RS232 port on removable, 4-pole terminal block
- Casing: ABS, water-tight housing, for wall installation with 4 screws
- Protection Rate: IP66
- Dimensions: 240 x 300 x h 120 mm; fixing template 220 x 185 mm
- Weight: approx. 1.5kg
- Analogic Inputs: precision better than 0.3% FS; repeatability better than 0.2% FS
- Temperature Input: removable, 2-pole terminal block, for connection of Pt100 sensor; standard range: -50…+200 °C
- Input for CLE Cell: available on 4-pole removable terminal block, for 2-electrode amperometric cell
(copper/platinum + shield)
- Power Supply: 230 V~ ±10% 50/60 Hz (115 or 24 V~, or 24 V- upon request)
- Power Consumption: 7 VA max (models with relay output)
70 VA (models with K1 configured as “dosing-shot”) - Protections: PTC fuse appropriate to the power supply voltage.
If K1=dosing-shot and 230 V~ power supply, F 250 mA, 5×20 fuse - Environment: storage temperature -20 to +60 °C
operating temperature -10 to +50 °C
RH max 90% noncondensing - Display: 2-row, alphanumeric LCD (16 digits per row), with backlight.
- Digital Inputs: 1 OFF/FLW input for disabling outputs; for voltage-free contact; 5 V voltage at terminals, current max 5 mA.
1 AUX (auxiliary) input, available on special models.
1 TC (temperature control) input, internal input, only for K1= dosing-shot - Dosing-shot Output: factory setting of K1 relay for directly driving of the pump magnet; standard pulse 300 V and 90 msec, max frequency 120 pulses/minute (not available for 24 V- or 24 V~ power supply versions)
- Current Output: 0-20 or 4-20 mA (selectable through software), max load 600 Ohm, error max 0.5% FS; provided with galvanic insulation.
- Serial Output: RS232 port on removable, 4-pole terminal block
- Casing: ABS, water-tight housing, for wall installation with 4 screws
- Protection Rate: IP66
- Dimensions: 240 x 300 x h 120 mm; fixing template 220 x 185 mm
- Weight: approx. 1.5kg
- Analogic Inputs: precision better than 0.3% FS; repeatability better than 0.2% FS
- Temperature Input: removable, 2-pole terminal block, for connection of Pt100 sensor; standard range: -50…+200 °C
- Input for CLE Cell: available on 4-pole removable terminal block, for 2-electrode amperometric cell
(copper/platinum + shield)
Vendor Information
-
-
Chemicals, Laboratory Reagents & Standards, Chemicals, Laboratory Chemicals
E. coli Chromogenic Medium