Radon Scout Plus
$4,772.40
RADON SCOUT Plus
MANUFACTURER: SARAD GmbH
REF NO: 11014030
- Description
- Size Guide
- SPECIFICATION
- DOCUMENT
- Vendor Info
- More Products
Description
Radon scout Plus
Description of radon scout Plus
The Radon Scout offers the highest senitivity of its class exceeding the sensitivity of its competitors many times. This results in the lowest detection limit and statistical error. The thousands of times proven and US-EPA certified instrument will show the time distribution of radon even at levels below 100 Bq/m³. Thus, meteorological influences as well as ventilation effects can be detected easily.
Due to its outstanding sensitivity, the instrument is perfectly suited for direct controlling of ventilation equipment by a potential-free switch contact. As an accessory, we offer a wireless switch which can be directly connected to the instrument. Thus, very low effort is required to control fans such as kitchen hoods, wall or window ventilators anywhere in your home. It is even possible to upgrade older instruments. The switch threshold for the Radon Scout is fixed to the EU target value of 300 Bq/m³.
Two standard D-size cells enable an autonomous operation over a few months. The radon detector Radon Scout PLUS offers an additional connection for an external power supply. In fact, this allows an operation over infinite time periods.
The size of the data memory was designed to get a high-resolution time distribution. Any number of measurement series may be created. The acquired data can be loaded by a PC even if a measurement is in progress.
The included Radon Vision software allows sophisticated data evaluation and handles remote connection as simple as a direct cable link.
Because of the high sensitivity, variations of the activity concentration will detected accurately even in case of low radon levels. The radon detector works in diffusion mode so that a possible influence of thoron can be excluded. The measurement chamber, equipped with a semiconductor detector and high voltage collection, is immune to ambient humidity.
Sensors for temperature and humidity (barometric pressure in addition in case of Radon Scout PLUS) are very useful add-ons. The integrated tilt detector will give a signal if the radon detector was moved from its original position during the measurement.
The instrument has been certified by the US-EPA/NRSB and other international authorities.
The instrument comes with a DAkkS compliant calibration certificate. The radon calibration process fulfills the requirements of the DIN ISO/IEC EN 17025:2018.
Applications:
- for long-term measurements of the activity concentrations of airborne radon (222 Rn) in homes, at workplaces (also underground)
- eminently suitable for the direct control of ventilation systems
Features:
- highest sensitivity in its class
- US EPA certified device
- shows the radon concentration time-resolved, even below 100 Bq/m³
- Radon Scout switching threshold set to 300 Bq/m³, freely selectable with Radon Scout Plus
- battery-powered, autonomous operation for several months, additional plug-in adapter for the Radon Scout Plus
- remote access and data transmission possible
- sensors for temperature, humidity and an integrated motion detector, the Radon Scout Plus also has an air pressure sensor
- LCD display on the Radon Scout Plus
- DAkkS-accredited calibration according to DIN EN ISO/IEC17025:2018
SPECIFICATION:
| Radon measurement | |
| Detector type | semiconductor detector with HV-chamber |
| Measuring range | 1 … 10.000.000 Bq/m³ |
| Sensitivity | 1,8 imp/min @ 1000 Bq/m³ (independent of humidity) |
| Accuracy | <=6% |
| Stat. error | 200 Bq/m³ with 20% stat. error (1) at 1h interval
1.000 Bq/m³ with stat. error (1) < 10 % at 1h interval 100 Bq/m³ with 17% stat. error (1) at 3 h interval |
| Response time | 120 minutes to 95% of final value |
| Integrated sensors | relative humidity, temperature, tilt and
Radon Scout Plus: barometric pressure |
| Integration interval | 1 h or 3 h selectable
Radon Scout Plus: 1…255 min adjustable per minute |
| Potential-free switch contact | threshold level fixed to 300 Bq/m³ (EU reference value) Radon Scout Plus: adjustable threshold max. 60 V, 400 mA, peak current 1.5A |
| Instrument | |||
| Data storage | non-volatile memory (circular structure) 672 data records
Radon Scout Plus: 16.383 data records |
||
| Power supply | 3,3 V, 2 x D-size cell (Ni-Cd, NiMH or Alkaline)
Radon Scout Plus: additional external DC adapter |
||
| ATEX category | No | ||
| Interfaces | USB, RS232 | ||
| Battery life | > 90 days | ||
| Operation | single switch, measurement/stand-by (lock-function)
Radon Scout Plus: display with backlight |
||
| Environmental conditions | temperature 0 … 40°C
rel. humidity 0 … 95%, non-condensing bar. pressure 800 … 1.100 hPa |
||
| Dimensions | 175 x 135 x 55 mm | ||
| Weight | 800 g (incl. batteries) |
| Radon Vision Software
(incl.) |
set-up, data download interactive graphical display (zoom, pan, fit data-cursor, marker for tilt and start of a new measurement, error bars, smoothing) selective ASCII export (EXCEL format) selective graphical protocol print (space for individual header, user comments) calculation of average concentration and exposure automatic created file names and path structure switch over between US and SI-units (Bq/m³ / pCi/L) |
| Scope of delivery | Radon Scout / Radon Scout Plus |
| Data storage | non-volatile memory (circular structure) 672 data records
Radon Scout Plus: 16.383 data records |
| Power supply | 3,3 V, 2 x D-size cell (Ni-Cd, NiMH or Alkaline)
Radon Scout Plus: additional external DC adapter |
| ATEX category | No |
| Interfaces | USB, RS232 |
| Battery life | > 90 days |
| Operation | single switch, measurement/stand-by (lock-function)
Radon Scout Plus: display with backlight |
| Environmental conditions | temperature 0 … 40°C
rel. humidity 0 … 95%, non-condensing bar. pressure 800 … 1.100 hPa |
| Dimensions | 175 x 135 x 55 mm |
Closer to your application
| Weight | 800 g (incl. batteries) |
| Radon Vision Software
(incl.) |
set-up, data download interactive graphical display (zoom, pan, fit data-cursor, marker for tilt and start of a new measurement, error bars, smoothing) selective ASCII export (EXCEL format) selective graphical protocol print (space for individual header, user comments) calculation of average concentration and exposure automatic created file names and path structure switch over between US and SI-units (Bq/m³ / pCi/L) |
| Scope of delivery | Radon Scout / Radon Scout Plus |