Smart Radon Sensor – CO2

$3,129.36

Smart radon sensor – CO2

MANUFACTURER: SARAD GmbH

REF NO: 11014011

 

 

 

 

 

 

 

 

SKU: EA-1371-408

Description

Smart Radon Sensor -CO2

Description of smart radon sensor

SARAD’s outstanding detector technology defines a new level of Radon monitoring in buildings and workplaces, as requested by the recently introduced EU radiation protection regulations. The Smart Radon Sensor offers very high sensitivity in a very small and shapely housing. Due to the detector technology, the instrument will not be affected by any variations in temperature and humidity, external radiation, mechanical shocks, and vibrations. The measurement starts as soon as the power is connected.
Various interfaces for system integration are available. The RS485 network interface can be operated either with the MODBUS RTU (industrial standard) or the proprietary SARAD protocol. The analogous output makes it easy to connect the Smart Radon Sensor to existing PLC’s. A stand-alone ventilation control system can be easily realized by the alert switch output. If no data line is available, the optionally integrated ZigBee wireless interface can be used for connection to a host system. The wide range of power input accepts all in building automation commonly used voltage supplies. For easy installation, all connections are available through a screw terminal. Two AAA batteries are integrated to continue the measurement in case of power interruptions. All measurement results will be stored locally in non-volatile circular memory. Sensors for coal dioxide (CO2)temperature, and humidity are integrated into the current version as a standard. There are options for barometric pressure measurement or without both CO2 and pressure sensors available.
The Smart Radon Sensor comes with the DAkkS-accredited calibration certificate according to the DIN EN ISO/IEC 17025:2018 and the latest release of the Radon Vision software. A standard calibration is performed at a radon concentration of 3 000 Bq/m³, with the prior determination of the background effect. A second calibration point at a radon concentration of 300 Bq/m³ is possible on demand and at additional cost. A separate calibration certificate is created for each calibration point.
We use only high-quality and instrument-grade components for this product, which is manufactured in our company’s location in Dresden/Germany.
Environmental conditions -10°C … 40°C

0% rH … 95% rH, non-condensing

800…1100 hPa

Power supply 12 V to 24 VDC

2 x AAA batteries only as backup (approx.10 days)

ATEX category no
Data storage non-volatile memory with circular structure for 16383 data records in any number of test series; saves all parameters including temperature and humidity each interval (from 1 to 255 minutes)
Interfaces RS485 (MODBUS RTU and SARAD protocol) ZigBee wireless (SARAD Net Monitors) – optional 4…20 mA analogous output (range configurable)  alert switch (potential free contacts, max 40 V, 250 mA, peak current 0.75A) all interfaces with screw terminals
Alert indication red, yellow and green LED indicators for alert and state of operation
Operation automatic start after powering up
Radon Vision Software (incl.)   set-up, data download, interactive graphical display (zoom, pan, fit data-   cursor, marker for tilt and start of a new measurement, error bars, smoothing) selective ASCII export (EXCEL format) selective graphical protocol print (space for individual header, user comments) calculation of average concentration / exposure automatic created file names and path structure   switch over between US / SI-units (Bq/m³/pCi/L)
 Dimensions 82mm x 96mm x 44mm
Weight 150 g incl. batteries
Scope of delivery bracket for wall mounting

USB cable

batteries 2 x AAA

user manual & Software (electronic version)

(1-  DAkks accredited calibration certificate according to DIN EN ISO/IEC 17025:2018  :(1- point calibration with 3000 Bq/m³) power supply adapter (on request