Temperature & Standard Signal Transducer, P20

$76.50

Manufacturer: LUMEL

Product Code: P20

SKU: AM-5148-28

Description

Temperature & Standard Signal Transducer, P20

Temperature & Standard Signal Transducer, P20

Description of Temperature & Standard Signal Transducer, P20

The P20 programmable transducer is destined to convert the temperature, resistance, voltage from shunt and standard signals into a standard constant-current or constant-voltage signal.

The transducer output signal is galvanically insulated from the input signal and the supply.

Features:

  • input: programmable universal input,
  • output: 0/4…20mA, 0…10V.

The transducer is fully configurable through the PD14 programmer and free LPConfig software.
By means of the programmer one can change

  • input type,
  • averaging time of the measurement,
  • rescale the analog output acc. to the individual output characteristic,
  • read out the measured value.

External dimensions: 22,5 x 120 x 100mm

SPECIFICATIONS

Basic parameters:

analog output galvanically isolated:

 

current

0/4… 20 mA

voltage

0…10 V

load resistance of the current output

≤ 500 W

load resistance of the voltage output

≥ 500 W

accuracy class¹

0.2

averaging time of the transducer:

 

range: d.c. current [mA], d.c. voltage [V]

≥ 0.1 s

other ranges

≥ 0.3 s

power consumption

< 3 VA

preheating time of the transducer

10 min

response time of the transducer:

 

range: d.c. current [mA], d.c. voltage [V]

≥ 0.2 s

other ranges

≥ 0.4 s

current flowing through RTD

< 0.2 mA

resistance of wires connecting RTD with the transducer

≤ 10 W

Rated operating conditions:

supply depending on the execution code

85… 253 V a.c./d.c.
20… 40 V a.c./d.c.

frequency of the supplya.c. voltage

45… 65 Hz

ambient temperature

– 20…23…55ºC

storage temperature

– 25…85ºC

related air humidity

< 95%
(condensation inadmissible)

working position

any

Input parameters:

resistance of voltage input [V]

> 1 MΩ

resistance of current input [mA]

12 Ω ±1%

Sustained overload:

TC and RTD

1.1 Xn

voltage, current and resistance

1.3 Xn

Short duration overload :

input voltage

5 Un

current input

10 In

Ensured protection level acc. to EN 60529:

housing

IP 40

electrical connections

IP 20

Weight

0.125 kg

Dimensions

22.5 x 120 x 100 mm

Fitting

on a 35 mm rail holder

Electromagnetic compatibility:

noise immunity

acc. to EN 61000-6-2

noise emission

acc. to EN 61000-6-4

Safety requirements acc. to EN 61010-1:

installation category

III

level of pollution

2

working voltages in relation to the ground:

 

supply

300 V²

input

50 V

output

50 V

altitude above sea level

< 2000 m

1) A part of sub-ranges for thermocouples and RTD has a specified individual class (see table 1)
2) Execution for supply voltage 230 V

Vendor Information