WDXRF SULFUR ANALYZER – SPECTROSCAN CLSW-D3
$72,261.96
MANUFACTURER: SPECTRON
MODEL: SPECTROSCAN CLSW-D3
ACCESSORIES
- Description
- Size Guide
- Vendor Info
- More Products
Description
WDXRF SULFUR ANALYZER – SPECTROSCAN CLSW-D3
Product Description of WDXRF SULFUR ANALYZER – SPECTROSCAN CLSW-D3
SPECTROSCAN CLSW wavelength dispersive X-ray fluorescent WDXRF analyzer of organic chlorides in oil is used for determination of organic chlorine content in oil at refinery factories, in research laboratories of oil Industry, oil-transport companies.
Routine work with the Analyzer does not require high skills of the staff and can be fulfilled by a laboratory assistant with no control from analysts
ANALYTICAL PARAMETERS
SPECTROSCAN CLSW is a desk top with inbuilt computer and printer. It consists of two blocks: Spectrometer and vacuum pump. The spectrometer has a closed cycle water cooling system. The measuring tract is in vacuum chamber, but the sample is at the ambient conditions.
| Analytical parameters | |
| Determined element | Cl (Chlorine) |
| Limit of detection for200 s | 0,5 mg/kg |
| Chlorine concentration range | from 0 mg/kg to 50 mg/kg |
| Type of Chlorine line detection | Crystal diffraction |
| Crystal analyzer | С(002) – pyrolitic carbon |
| X-ray tube anode | Cr |
| Time for sample measurement | from 8 minutes |
| Technical specification | |
| Sample changer | Automatic lateral loading for 3 sample cups |
| Sample cups, diameter, volume |
∅32 mm, V 8 cm3 |
| ∅32 mm, V 8 cm3, ventilated | |
| X-ray tube capacity | less 200 W |
| Interface | display, thermo printer |
| Dimensions and weight (not more) | 530х480х340 mm, 40 kg– spectrometer |
| 330х230х380 mm, 9 kg – vacuumpump | |
| Power supply | 220 V, ~ 50 Hz, 750 W |
ADVANTAGES
Due to the original design no He blowdown is required, providing LLD of 1 ppm
- Simple in operation;
- Desk top instrument;
- The sample data and analysis results are shown in display and printed out by inbuilt printer;
- Special ventilated cups are developed for volatile petroleum products;
- 3-position sample automatic changer makes analysis of sample replicateseasy.
Due to lateral position of the sample:
- Error due to water and air bulbs in petroleum products are excluded;
- Contamination with petroleum products of the inner parts of the analyzer is excluded;
- Additional errors due to contamination of the additional protection film is excluded;
- Sample loading unit is easily cleaned.
Minimum operator actions:
- Sample preparation;
- Put in the number or name of the sample from inbuilt key-board;
- Fill in two sample cups with the sample;
- Put in the samples into the analyzer and start measurements.
All the rest operations are run automatically.
APPLICATIONS